Gester International Co., Ltd

Gester International Co., Ltd

Manufacturer from China
Active Member
5 Years
Home / Products / Mask Test Machine /

GESTER 20 Degree ISO 13982 Masks Leakage Tester

Contact Now
Gester International Co., Ltd
City:quanzhou
Province/State:fujian
Country/Region:china
Contact Person:Mrsjoyce
Contact Now

GESTER 20 Degree ISO 13982 Masks Leakage Tester

Ask Latest Price
Brand Name :GESTER
Model Number :GT-RA08A
Certification :CE, ISO
Place of Origin :China
MOQ :1 set
Payment Terms :L/C, D/A, D/P, T/T, Western Union, MoneyGram
Supply Ability :200 sets/month
Delivery Time :10-30 days
Packaging Details :wooden packaging
Keyword :Protective Clothing and Masks Leakage Tester
Type :Mask Test Equipment
Working temperature :(20±5)℃
Humidity :≤60%
Standards :EN149, ISO 13982
After-sales :installation and training service
more
Contact Now

Add to Cart

Find Similar Videos
View Product Description

Protective Clothing and Masks Leakage Tester GT-RA08A

Product Introduction


Application:

To test protective clothing against solid airborne particles including radioactive contamination.

The machine use “inward leakage in %”, ratio of the concentration of contaminant in the ambient atmosphere to the concentration of the contaminant in the suit, to evaluate the quality of protective clothing.

Standards:
GB/T 29511-2013 Protective clothing- Chemical protective clothing against solid particulates4.3.1, Annex A
GB 2626-2019Respiratory protection—Non-powered air-purifying particle respirator 6.4
GB 19083-2010Technical requirements for protective face mask for medical use ANNEX B
ISO 13982,
BS EN ISO 13982 ,
EN149,
EN 1073-1-2016

Key Specification

Model

GT-RA08A

Environmental Requirement

Working temperature

(20±5)℃

Humidity

≤60%

Used space

≥4000×5000×3000mm,

water in/drain pipe, room with good ventilation

Walked-in Testing Room

Inner effective size

1800mm×1500mm×2100mm(L×W×H)

Outer size of testing room

1950mm×1650mm×2500mm(L×W×H)

Control cabinet

700mm×700mm×1600mm(L×W×H)

Environmental Analog detection system

Particle aerosol concentration

(10±1) mg/mPP3

Aerodynamics size distribution of particle

(0.02~2)μm

Mass median diameter

About 0. 6μm

Concentration variation in effective space

≤10%

Occurrence amount

≥100L/min

Particle Concentration Detector

2 sets

Dynamic range

(0.001~200)mg/m3

accuracy

±1%

response time

≤500 ms

Horizontal Pedal Drive Test-bed

Running speed

(5±0.5)km/h

Other Accessories

Air duct circulation and maintenance of cleaning equipment

Feature


A standard aerosol of sodium chloride particles is generated inside a test chamber in which a test subject, wearing the protective suit under test, carries out a predetermined sequence of test exercises. The inward leakage at each sampling position inside the suit is measured by means of flame photometry.We use following indicator to judge the quality of protective clothing.
- Lijimn: inward leakage for a given test subject (i),suit(j),exercise (m) and sampling position (n)
- Cijmn: concentration of aerosol measured at the sampling point inside the suit for a give test subject(i), suit(j(, exercise(m) and sampling position(n)
- LS: total inward leakage per suit (average over all exercises and sampling positions)
- LH: total inward leakage per human test subject (average over all exercises, sampling positions and suits worn by that test subject)
- LE: total inward leakage per exercise (average over all suits and sampling positions)
- Lp: total inward leakage per sampling position (average over all suits and exercises)
- LEP: total inward leakage per sampling position and per exercise (average over all suits)

- : mean total inward leakage (average over all test subjects, suits, exercises and sampling positions)

Controlling & Analysis system

1.Protective Clothing Against Solid Airborne Particles Tester Central control console, computer programmed control, prompting experiment process;

2.Monitoring test data automatically, calculating individual test inward leakage rate, total inward leakage rate, average inward leakage automatically, output testing result in various way.

GESTER 20 Degree ISO 13982 Masks Leakage Tester

Inquiry Cart 0